Accelerometer Calibration Services
Details...

  • A2LA Accredited to ISO/IEC Guide 17025 -  2005 through December 31, 2007 by virtue of the onsite assessment of the calibration laboratory and evaluation of all activity related to the assessment by the A2LA Accreditation Council. Scope and A2LA Accreditation Certificate numbered 1644.01.
  • Calibrations performed on accelerometers from other manufacturers
  • Traceable to NIST in accordance with ANSI/NCSL Z540-1-1994 standards
  • Ability to calibrate piezoelectric, peizoresistive, variable capacitance and integral electronics type accelerometers
  • Average turn-around time of 5 to 10 working days

    NOTE: Robert A. Denton, Inc. does not repair accelerometers. Accelerometers must be returned to the original manufacturer for repair.

     

    Calibration Protocol: Accelerometer calibrations are normally performed at a 10 g level from 20 Hz through the usable frequency range up to 50kHz. A resonance search is also provided up to 50kHz. Robert A. Denton, Inc. performs DS415 Service (sensitivity, 10 g, 100 hz, Freq Resp Fn) for uniaxial and triaxial accelerometers. In addition, the DS270 Shunt Calibration is available in conjunction with the DS415 calibrations for piezoresistive type accelerometers.

    Can't find what you are looking for? Please contact our Customer Service department for assistance.